The design and construction of a photoelectric spectrophotometer for analytical research

dc.contributor.author Howe, Donald
dc.contributor.department Chemistry
dc.date 2018-08-22T22:56:53.000
dc.date.accessioned 2020-06-30T07:25:18Z
dc.date.available 2020-06-30T07:25:18Z
dc.date.copyright Thu Jan 01 00:00:00 UTC 1942
dc.date.issued 1942
dc.description.abstract <p>A spectrophotometer has been built, consisting of three parts, a monochromator, a photocell amplifier, and a recorder;The monochromator used a new optical system, a modification of the Wadsworth mounting for concave reflection diffraction gratings;Upon calibration of the monochromator using a helium spectrum tube as a light source, the wave length scale was found to be linear;Seven separate amplifiers were constructed, of which one was finally used. It suffered less from random currents and drift than did the other six, but was still so erratic that results could be reproduced with a precision of only five percent;A recorder mechanism was constructed, and was found to be too unreliable for good analytical work even after a number of changes were made in the circuit;The transmittancies were obtained for two glass filters calibrated by the Bureau of Standards. These were found to check reasonably well with the calibrated values, except at the extreme ends of the spectrum. This comparison afforded a qualitative estimation of the amount of stray light present in the monochromator. A barrier layer cell coupled directly to a galvanometer was used in these determinations;Results of high precision were found to be not obtainable in the instrument constructed without the use of optical and electrical parts of precision manufacture.</p>
dc.format.mimetype application/pdf
dc.identifier archive/lib.dr.iastate.edu/rtd/12988/
dc.identifier.articleid 13987
dc.identifier.contextkey 6866714
dc.identifier.doi https://doi.org/10.31274/rtd-180813-14351
dc.identifier.s3bucket isulib-bepress-aws-west
dc.identifier.submissionpath rtd/12988
dc.identifier.uri https://dr.lib.iastate.edu/handle/20.500.12876/66416
dc.language.iso en
dc.source.bitstream archive/lib.dr.iastate.edu/rtd/12988/r_DP12106.pdf|||Fri Jan 14 19:34:36 UTC 2022
dc.subject.disciplines Analytical Chemistry
dc.subject.keywords Spectrophotometer
dc.subject.keywords Photoelectricity
dc.subject.keywords Chemistry
dc.subject.keywords Analytic
dc.subject.keywords Analytical Chemistry
dc.title The design and construction of a photoelectric spectrophotometer for analytical research
dc.type article
dc.type.genre dissertation
dspace.entity.type Publication
relation.isOrgUnitOfPublication 42864f6e-7a3d-4be3-8b5a-0ae3c3830a11
thesis.degree.level dissertation
thesis.degree.name Doctor of Philosophy
File
Original bundle
Now showing 1 - 1 of 1
No Thumbnail Available
Name:
r_DP12106.pdf
Size:
4.53 MB
Format:
Adobe Portable Document Format
Description: