Three‐dimensional probe reconstruction for atomic force microscopy
Three‐dimensional probe reconstruction for atomic force microscopy
Date
1994-07-01
Authors
Vesenka, James
Miller, Richard
Henderson, Eric
Henderson, Eric
Miller, Richard
Henderson, Eric
Henderson, Eric
Major Professor
Advisor
Committee Member
Journal Title
Journal ISSN
Volume Title
Publisher
Altmetrics
Authors
Henderson, Eric
Person
Research Projects
Organizational Units
Journal Issue
Series
Department
Zoology and GeneticsMathematics
Abstract
Colloidal gold particles are used as hard, spherical imaging targets to assist in the three‐dimensional reconstruction of the atomic force probe apex. Probe reconstructions are shown to be accurate to 1 nm resolution and dynamically change as the sample is scanned, emphasizing the utility of colloidal gold particles as in situ calibration standards for image reconstruction of a coadsorbed specimen.