Equivalency of near infrared transmission instruments for grain analyzers

dc.contributor.advisor Charles Hurburgh
dc.contributor.author McGinnis, Samantha
dc.contributor.department Department of Agricultural and Biosystems Engineering (ENG)
dc.date 2018-08-11T12:18:04.000
dc.date.accessioned 2020-06-30T03:07:48Z
dc.date.available 2020-06-30T03:07:48Z
dc.date.copyright Fri Jan 01 00:00:00 UTC 2016
dc.date.embargo 2001-01-01
dc.date.issued 2016-01-01
dc.description.abstract <p>The Grain Inspection Packers and Stockyard Administration (GIPSA) has one approved near infrared transmission (NIRT) instrument for the official inspection of grains. This is believed to be more accurate multiple makes and models used. This study focused on determining if more than one make and model of transmission instruments can be used. Three National Type Evaluation Program (NTEP) approved NIRT instruments coded A, B, and C were evaluated for equivalence. To be equivalent, the variation of results of all the instruments would be less than or equal to that of one instrument make and model on a population of samples. The study used 5 copies of each instrument, for a total of 15 machines. The number of samples used were, 250 wheat, 100 barley, 145 soybeans, and 149 corn. The samples were passed through all 15 machines three times each. Results for wheat (protein), barley (protein), soybean (protein and oil), and corn (protein and oil) were collected using the NTEP approved calibrations for each unit, with and without slope-bias standardization. A least mean squares analysis partitioned the variance by A-A, A-B, B-B, B-C, and C-C. The instruments as set up were not equivalent because A-B, A-C, and B-C were significantly larger than the within brand comparisons A-A, B-B, and C-C (p<0.0001).</p>
dc.format.mimetype application/pdf
dc.identifier archive/lib.dr.iastate.edu/etd/15974/
dc.identifier.articleid 6981
dc.identifier.contextkey 11169450
dc.identifier.doi https://doi.org/10.31274/etd-180810-5601
dc.identifier.s3bucket isulib-bepress-aws-west
dc.identifier.submissionpath etd/15974
dc.identifier.uri https://dr.lib.iastate.edu/handle/20.500.12876/30157
dc.language.iso en
dc.source.bitstream archive/lib.dr.iastate.edu/etd/15974/McGinnis_iastate_0097M_16032.pdf|||Fri Jan 14 20:49:22 UTC 2022
dc.subject.disciplines Agriculture
dc.subject.disciplines Bioresource and Agricultural Engineering
dc.subject.keywords Equivalence
dc.subject.keywords grain
dc.subject.keywords NIRT
dc.title Equivalency of near infrared transmission instruments for grain analyzers
dc.type thesis en_US
dc.type.genre thesis en_US
dspace.entity.type Publication
relation.isOrgUnitOfPublication 8eb24241-0d92-4baf-ae75-08f716d30801
thesis.degree.discipline Agricultural and Biosystems Engineering
thesis.degree.level thesis
thesis.degree.name Master of Science
File
Original bundle
Now showing 1 - 1 of 1
No Thumbnail Available
Name:
McGinnis_iastate_0097M_16032.pdf
Size:
630.33 KB
Format:
Adobe Portable Document Format
Description: