A beta-ray spectrometer for coincidence measurements

Date
1953-05-15
Authors
Nichols, R.
Pohm, A.
Talboy, J.
Jensen, E.
Journal Title
Journal ISSN
Volume Title
Publisher
Altmetrics
Authors
Research Projects
Organizational Units
Ames Laboratory
Organizational Unit
Journal Issue
Series
Abstract

An intermediate image beta-ray spectrometer has been constructed utilizing the focusing properties of a U-shaped magnetic field distribution. The construction and performance of the instrument are discussed. A transmission of ten percent has been attained with a resolution of about six percent. A scintillation spectrometer is used in conjunction with the intermediate image spectrometer for the purpose of making coincidence measurements.