An instrument for internal standard flame photometry and its application to the determination of calcium in the rare earths
Date
1950-07-25
Authors
Heidel, Robert
Fassel, Velmer
Major Professor
Advisor
Committee Member
Journal Title
Journal ISSN
Volume Title
Publisher
Abstract
Details are given for simple modifications of a laboratory monochromator so that selected internal standard line pairs from flame spectra may be isolated and measured simultaneoualy with a multiplier phototube differential photometer. The basic design is applicable to most spectrometers or spectrographs. An atomizing system with provisions for sample recycling and recovery is also described. This equipment has been applied to the determination of calcium in the rare earths in the concentration range of 0.025 to 2.5%. Average deviation for a series of repeated determinations at various concentrations is about plus or minus 1.5%.
Series Number
Journal Issue
Is Version Of
Versions
Series
ISC-107
Academic or Administrative Unit
Type
article