Use of AliveCor Heart Monitor for Heart Rate and Rhythm Evaluation in Dairy Water Buffalo Calves (Bubalis Bubalis)
Date
Authors
Major Professor
Advisor
Committee Member
Journal Title
Journal ISSN
Volume Title
Publisher
Abstract
Smartphone-based ECG recording devices are a promising new diagnostic in veterinary medicine. However, these devices have not yet been demonstrated to work in water buffalo. A commercial, non-invasive, smartphone-based ECG (AliveCor Vet TM) monitoring device was field tested on water buffalo calves. Six neonatal water buffalo calves had an ECG simultaneously recorded with a standard medical ECG recorder and the AliveCor Vet device. A correlation of 0.75053 was observed between average heart rates obtained between AliveCor recordings when compared to those from the 6 lead ECG. 100% Rhythm agreement was met for all 12 recordings. The recordings of 4/6 calves most closely resembled lead aVR, recordings of 2/6 calves more closely resembled lead II. No significant difference was observed between standard and AliveCor quality scores (P=0.3632). We conclude that the AliveCor device allows for accurate rate diagnosis, and accurate diagnosis of a normal sinus rhythm in neonatal diary water buffalo calves in a field setting.
Series Number
Journal Issue
Is Version Of
Versions
Series
Academic or Administrative Unit
Type
Comments
This article is published as Smith J, Ward J, Urbano T, Mueller M (2016). Use of AliveCor heart monitor for rate and rhythm evaluation in dairy water buffalo calves (Bubalis bubalis). J Dairy Vet Anim Res 4(2):00113. doi:10.15406/jdvar.2016.04.00113. Posted with permission.