The AliveCor KardiaMobile ECG device allows electrocardiogram assessment in awake bonobos (Pan paniscus)
Date
2023-04-04
Authors
Olds, June E.
Goldacker, Ashley
Huneycutt, David
Huneycutt, Sarah P.
Taglialatela, Jared P.
Major Professor
Advisor
Committee Member
Journal Title
Journal ISSN
Volume Title
Publisher
American Veterinary Medical Association
Abstract
OBJECTIVE
To determine the diagnostic utility of a smartphone-based ECG device (Alivecor KardiaMobile) in awake bonobos (Pan paniscus).
ANIMALS
7 adult bonobos in human care.
PROCEDURES
Bonobos were trained with positive reinforcement to hold 1 finger from each hand onto the KardiaMobile sensors for 30 seconds to obtain an ECG reading. Ten ECG tracings were recorded from each bonobo and evaluated by a veterinarian, a veterinary cardiologist, and a human cardiologist for tracing quality, tracing length, heart rate, identification of P-waves, and presence and quantification of premature ventricular or atrial contractions.
RESULTS
6 of the 7 bonobos were trained within 21 weeks to allow the collection of 10 diagnostic quality ECG tracings. The average heart rate recorded was 87 bpm (range = 60 to 118 bpm). Potential abnormalities identified by the KardiaMobile included premature ventricular contractions in 2 male bonobos and 1 premature atrial contraction in another male. There was strong agreement by reviewers in all ECG parameters except for the identification of P-waves.
CLINICAL RELEVANCE
The Alivecor KardiaMobile device has diagnostic utility as a screening tool for use in bonobos in human care. The training was accomplished to yield diagnostic ECG readings of acceptable duration in awake bonobos. Given the prevalence of cardiovascular disease in great apes, this technology may identify a subset of great apes who may benefit from early intervention and treatment in an effort to delay the progression of cardiac disease.
Series Number
Journal Issue
Is Version Of
Versions
Series
Academic or Administrative Unit
Type
article
Comments
This article is published as Olds, June E., Ashley Goldacker, David Huneycutt, Sarah P. Huneycutt, Jared P. Taglialatela, and Jessica L. Ward. "The AliveCor KardiaMobile ECG device allows electrocardiogram assessment in awake bonobos (Pan paniscus)." American Journal of Veterinary Research (2023): 1-7.
DOI: 10.2460/ajvr.23.01.0013.
Copyright 2023 American Veterinary Medical Association.
Attribution-NonCommercial 4.0 International (CC BY-NC 4.0).
Posted with permission.